Strong technology
The integrated collaborative measurement of mass concentration and heavy metal concentration of airborne particulate matter can analyze the distribution of components in airborne particulate matter
Atmospheric heavy metal on-line analyzer is a dual function fully automatic on-line analysis instrument, which combines X-ray fluorescence (XRF) non-destructive testing technology, beta ray absorption testing technology and air particulate matter automatic enrichment technology, and conforms to the American EPA io-3.3 standard method. It can not only monitor the mass concentration of air particles, but also quantitatively analyze the elements in the particles.
The instrument has a detection limit of 10pg / m3, which can be widely used in air quality monitoring, pollution tracing and source analysis, environmental assessment and other fields.
The integrated collaborative measurement of mass concentration and heavy metal concentration of airborne particulate matter can analyze the distribution of components in airborne particulate matter
Equipped with a variety of particle cutters for users to choose, the distribution of heavy metal particles in different dynamic diameters can be measured
The detection sensitivity of 30 heavy metals such as Pb, CD, Cr and as is pg / m3, and the element types can be expanded
The main machine adopts standard 19 "chassis, which can be installed on both mobile monitoring vehicle and standard cabinet of automatic air station
20 years of XRF technology accumulation, lightning protection industrial grade electronic and electrical design, the instrument has high operation reliability and stability, 24 hours / 365 days trouble free operation, but also has intelligent fault diagnosis and self troubleshooting function
No stray peak optical path design, built-in multi band filter, automatic switching to achieve low background measurement
Built in calibration module, XRF spectrum abscissa and ordinate can be automatically calibrated at the same time, the calibration frequency can be set arbitrarily, and a specific metal internal standard source is built in, which can realize automatic quality control and alarm
Through the mass flow controller and temperature and pressure sensors, the sampling flow can be stably controlled, and the automatic flow calibration can be realized
With RS232 / 485, USB, GPRS, ADSL, Ethernet and other data transmission modes, it can realize the remote transmission, monitoring and inspection of operation parameters, status and data
The system software adopts advanced hierarchical modular design, with powerful and flexible data setting, calibration, saving, display, query, analysis, transmission and other functions. The data conforms to the data transmission standard of online automatic monitoring (monitoring) system for pollution sources (HJ / T 212-2005). At the same time, users can quickly and easily customize and expand any other custom functions
Professional multiple processing, no radiation leakage of the instrument
Monitoring index | Tsp / PM10 / PM2.5 mass concentration, and can simultaneously monitor Pb (lead), CD (cadmium), Hg (mercury), as (arsenic), Cr (chromium), Cu (copper), Zn (zinc), Ni (nickel), BA (barium), Fe (iron), Ag (silver), Se (selenium), Br (bromine), te (tellurium), Sb (antimony), Sn (TIN), Ti (titanium), CO (cobalt), Mn (manganese), Pd (palladium), SC (scandium), etc Mo (molybdenum), V (vanadium), CS (cesium), GA (gallium), Ca (calcium), K (potassium), Sr (strontium), Cl (chlorine) and other 30 elements (other elements can be expanded according to the subsequent needs of users) |
Applicable standards |
HM:US EPA IO-3.3 Standard method《Determination of heavy metals in environmental particles by X-ray fluorescence spectrometry》 PM:《Technical requirements and detection methods for continuous monitoring system of ambient air particles (PM10 and PM2.5)》 |
Measuring range |
HM:0~100 μg/m3 PM:0~10 mg/m3 |
Low detection limit |
HM:10 pg / m3 (sampling time 4 hours, flow rate 16.7 L / min) PM:<0.5 μg/m3(24-hour average) |
Accuracy | HM: & lt; 15% (Pb, compared with the national standard laboratory method, the sampling time is more than 1 hour) < br / > PM: ± 5 μ g / m3 (within 24 hours) |
Protect |
M: Multiple X-ray radiation protection device, the radiation dose is less than 2.5 μ Gy / h (5 cm away from the chassis); the X-ray tube is automatically closed when the instrument door is opened to protect personnel PM: the ionizing radiation inside the instrument is close to the air background value |
Operation mode | Continuous automatic monitoring, a year of trouble free operation, and fault self recovery function |
Control mode | Computer control, self diagnosis and setting function, real-time state monitoring function, and can store more than ten years of monitoring data |
Operation interface | Chinese or English operation interface, touch screen operation, display real-time sampling flow, sampling time, measurement status, heavy metal concentration value (standard status or non-standard status can be selected), content curve and other information |
Installation method | Installation of standard cabinet in fixed station building or mobile vehicle mounted installation |
Ehm-6618 can continuously measure and analyze the mass concentration of various particles and heavy metals in the air, and has the following applications: 1. Air quality monitoring 2. Characteristic analysis of air particles 3. Air background measurement 4. Environmental assessment and licensing 5. Pollution source location and traceability 6. Pollution prediction and early warning
Gc-ms3200 gas chromatography m...
Gc-ms3200 gas chromatography m...
Gc-ms3200 gas chromatography m...
Gc-ms3200 gas chromatography m...
Gc-ms3200 gas chromatography m...
Fax:0512-66326763-801
E-mail:cheng@boxunyq.com
Website:http://www.lehjax.cn
Add:39 wangmi street, high tech Zone, Suzhou City, Jiangsu Province